CHAMBER DESIGN
As per GMP* guidlines for long duration test.
Control Accuracy of temperature & humidity.
Uniformity of conditions through out the chamber.
Direct display of humidity in % RH.
Pt 100 sensor for dry / wet bulb or capacitance type humidity
sensor available.
Printer interface compatibility.
PC interface with RS 232-C / or RS-485 for multiple networking of
controller.
Built in high / low conditions alarm.
Low water level cut off device.
Complete stainless steel chamber also available.
CONSTRUCTION :
Inner S. Steel heavy gauge, outer CRCA powder and coated finish.
PUF in-situ foam about 3" thick.
Viewing window (Std) / full view inner door. ( To select while
ordering. )
Castor Wheel for easy movability.
High efficiency Refrigeration system (CFC Free)*
Humidity Injection System with low water level cut off.
Microprocessor based controller with PID action for Dry/Wet,
Direc % RH, safety alarm.
SAFETY FEATURES :
Control system with inbuilt temp. deviation audio video alarm.
A wet heater safety device.
A dry heater safety device.
Individual HRC fuses for Compressor, Air heater, Wet Heater and
Mains.
Validation : HMH Offers Validation Protocols for DQ, IQ, OQ & PQ
Documentation complete with reports, calibration and test
certificates.
| TYPE
OF CONTROLLER : |
Order Code |
| mP
PID Controller with direct % RH setting (Pt 100 Sensor for
Dry / Wet or Capacitance sensor.) |
HETC_S |
| mP
PID Controller - do - but with Printer Interface ( centronic
port ) |
HETC_P |
| mP
P PID Controller - do - but with
PC Interface through RS 232-C/RS 485 for multiple networking |
HETC_PC |
| mP
PID Profile Controller for Ramp / Soak for cyclic
applications. |
HETC_PRO |
|
|
| TECHNICAL
SPECIFICATIONS : |
Working
Size |
Material
of Construction : Inside S.S. & Outside S.S.
Inside S.S. & Outside GI Powder Coat Finish. |
50
X 50 X 50 X Cm. Ht. |
| Humidity
Range : 40% RH to 98% RH |
50
X 50 X 70 X Cm. Ht. |
| Temp.
Range # : 100C to 600C |
60
X 60 X 90 X Cm. Ht. |
| Accuracy
Temp. / RH : + 0.20C & + 3% RH |
1m3 |
| ( # Low temp.
Range : -250C / -150C / -100C &
00C & 35% RH on request.) |
(Bigger
size on request) |
NOTE :
An economical model is available with PID controller with direct
% RH display having all above features but with out printer
interface. (*Optional). Due to continues R & D
model & specifications may change. Picture shown with
Optional accessories. Store unit in AC Rooms.
|